Your DFM is in calibration mode. To return to normal measurement mode try this.
1. Set the RUN/PROGRAM switch to program.
2. Toggle(push up then release) the option switch six times to reach CALIBRATION.
3. Toggle the SEND and TARE switches down at the same time to turn calibration off.
4. Set the RUN/PROGRAM switch to RUN.